CONTACT OUR SALES

Overview of Regulatory Requirement for Post-Marketing Medical Device Reporting in the USA and the EU

Overview of Regulatory Requirement for Post-Marketing Medical Device Reporting in the USA and the EU

Meteriovigilance is the term used in the context of surveillance of medical devices and its purpose is to improve the protection of health and safety of patients, healthcare professionals, and other users by reducing the likelihood of reoccurrence of incidents related to the use of a medical device.

Presented below in brief is the post-marketing regulatory landscape as exist for USFDA and EMA in terms of Medical Device Reporting.

  1. Regulations that Govern Medical Device Reporting

  • USFDA
  • 21 CFR PART 803
  • 21 CFR PART 806
  • EMA
  • Regulation (EU) 2017/745
  1. Type of Report and Reporting Timelines

  • USFDA
  • 5-day report: No later than 5 workdays after becoming aware of a reportable event that requires
  • Remedial action to prevent an unreasonable risk substantial harm to the public health.
  • A reportable event for which FDA made a written request.
  • 30-Day report: No later than 30 calendar days after becoming aware of a reportable death, serious injury, or malfunction
  • EMA
  • Serious public health threat: Immediately, but not later than

days after the manufacturer becomes aware of that threat.

  • Death or an unanticipated serious deterioration in a person’s state of health: Immediately, but not later than 10 days after the date on which the manufacturer becomes aware of the serious incident.
  • Other serious incident: Immediately, but no later than 15 days after manufacturer become aware of the incident.
  1. Method to submit Reportable Medical Device Reports

  • USFDA
    • Electronic Submissions Gateway (Web Trader)
    • Gateway to Gateway
  • EMA
    • Submission of Manufacturer’s Incident Report (MIR) form via email to the approving Competent Authority.
    • Under development: EUDAMED
  1. Requirement for submission of Periodic Report

  • USFDA
    • Not mentioned*
    • Guidance for Combination product has specific requirements; will be detailed in subsequent posting.
  • EMA
    • Class IIb and class III devices: Annual PSUR Submission
    • Class IIa devices: PSUR to be submitted at least every two years
  1. Requirement for Trend reporting

  • USFDA
  • Not mentioned
  • EMA
  • Manufacturers shall electronically report any statistically significant increase in the frequency or severity of incidents that are not serious incidents or that are expected undesirable side effects that could have a significant impact on the benefit-risk analysis, and which have led or may lead to risks to the health or safety of patients, users or other persons that are unacceptable when weighed against the intended benefits

References

  1. 21 CFR Part 803 — Medical Device Reporting
  2. 21 CFR Part 806 — Medical Devices; Reports of corrections and removals
  3. https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-reportmedical-device-problems (assessed on 17-Oct-2023)
  4. REGULATION (EU) 2017/745: Medical Devices Regulation https://www.ema.europa.eu/en/human -regulatory/overview/medical-devices (assessed on 17-Oct-2023)

 

 

About Soterius

Soterius is a strong team of pharma professionals who design customized, innovative, and cost-efficient processes for clinical safety, pharmacovigilance, and medical affairs. Our deep industry knowledge and up to date insights let us combine agile, people powered intelligence in pioneering customer centric solutions. Our innovative technology solutions include engagement tools and communications platforms to create a unified and compliant medical access facility. With a strong global presence, we provide comprehensive clinical and post marketed safety services, that include aggregate report writing, signal detection and management, global literature surveillance, risk management, case processing and regulatory reporting. We use state-of-the-art technologies to solve complex safety operations problems, be it case processing, intake, site reporting for clinical trials, or literature search and management. We have one of the most accurate solutions for case intake and case processing using AI.

We support companies from the initial development stage of a drug/vaccine to the approval and ultimate marketing of the therapy, supporting ongoing operations and regulatory commitments globally.

Authors

Disha Dangwal

Disha Dangwal

Manager- Clinical Safety and Pharmacovigilance

Disha is a pharmacist with 9+ years of pharmacovigilance experience. She is skilled in processing and review of spontaneous, clinical trial, global literature, E2B, medical device, and cosmetic reports. She has experience in reporting of cases to the US, Canada, EU, UK, and Australia. She has managed various projects and participated in several client audits and regulatory inspections as a lead representative and Subject Matter Expert.

Disclaimer

Copyright 2025 by Soterius, Inc. All rights reserved. Soterius logo are trademarks or registered trademarks of Soterius in all jurisdictions. Other marks may be trademarks or registered trademarks of their respective owners. The information you see, hear or read on the pages within this presentation, as well as the presentation’s form and substance, are subject to copyright protection. In no event, may you use, distribute, copy, reproduce, modify, distort, or transmit the information or any of its elements, such as text, images or concepts, without the prior written permission of Soterius. No license or right pertaining to any of these trademarks shall be granted without the written permission of Soterius (and any of its global offices and/or affiliates). Soterius reserves the right to legally enforce any infringement of its intellectual property, copyright and trademark rights. Any content presented herewith should only be considered for general informational purposes and should not be considered as specific to the requirements of any particular organisation or for any specific purpose. Soterius does not make any representations or warranties about the completeness, reliability, appropriateness, relevance, or accuracy of the content presented here.